Loading files
This file is pinned to the selected release. Download the original to open it in your design tool.
# BOM notes
- Rows 1 to 15 are numbered to match the callouts in `media/exploded.png`. Rows 16 to 18 are not numbered in the exploded view; the gas lines of row 16 are modelled as 6 mm tubes.
- Every line is priced. All prices are indicative USD figures for a single prototype in 2026, not quotes. Items 6, 9 and 12 are estimates for generic parts and carry the largest uncertainty. The branded 12 W fuel cell list price ($576) and the MSE 300 mL/min single-cell stack price ($1,396.95) were checked on 2026-09-25.
- Totals (checked by `docs/04-calcs/sizing.py`, HBN-CAL-001 v0.3 section 10): **$885** for all 18 lines; $820 if the school supplies the bench power supply (line 4); $800 if the bench socket is also already RCD protected (line 18).
- `budget_usd` is $885 (raised from $450 to $850 by Amish's decision of 2026-09-25, and to $885 on 2026-09-26 to cover the priced BOM, HBN-DDR-002), with the bench supply in the kit. The full kit meets it with no margin; without the supply it is $65 under.
- The H2Guard detector, fan and interlock (line 15) is costed in the H2Guard project and is **not** included in the total. The bench must not run without it.
- Line 10's vent valve is a needle valve set so that venting 300 to 20 kPa gauge takes 3 min or more (HBN-CAL-001 section 7). Its relief valve is set to 325 kPa gauge, and it carries a pressure switch that opens at 310 kPa gauge; the switch drives the relay in line 14, which breaks the electrolyzer supply (HBN-DDR-002). These added $20 to lines 10 and 14.
- Line 18 (plug-in 30 mA RCD) was added at TRL 3 because the safety sections require RCD protection and no line covered it.
- The existing lab table shown in the renders is not part of the BOM.
f14b49466This file is pinned to the selected release. Download the original to open it in your design tool.