Aerosol-Reducing Mask for Non-Invasive Ventilation
Loading README and project media…
November 27, 2020
Effective November 27 (2020), the ARM team is suspending any sales or distribution of the current version of the ARM.
Review of clinical trial data that became available to the team on November 26 shows that the ARM device is not as effective at reducing leak and aerosolization as earlier data had suggested. The preliminary data and analysis report is posted to the repository.
The clinical trial was not a requirement for regulatory approval from Health Canada and the device is still approved under an Interim Order. However the trial data suggests that while safe and effective at providing respiratory support, the device fails to meets its primary indication (protection of health care providers and other patients from infectious aerosols), we are withholding all sales and deployments of the device.
We are currently undertaking a detailed analysis of the device and causes of failure in order to create a road map for required modifications for a subsequent version. Such modifications and the necessary testing will likely not be completed in sufficient time to meet current urgent needs. All analysis reports and other documents will continue to be posted on the public repository.
The Minimizing Aerosolization In Non-invasive (MAIN) Ventilation COVID-19 mask is a sealed, filtered, full-face mask minimizing virus aerosolization and enabling the use of Non-Invasive Positive Pressure Ventilation (NIPPV) machines for patients with COVID-19, or similar symptoms. The mask adaptor is the primary connection between the sealed facemask and standard medical 22mm respiratory tubing, fittings and filters. An additive manufacturing method, also known as 3D printing, was used to produce the mask adapter because of its relative speed to make objects of complicated geometries and availability. Due to the medical nature of the adapter, certain specifications must be followed in its production to prevent any infectious transmission from the patient to health care workers and other patients.
Additional Slicer Settings:

Figure 1: Orientation of Adapter v1.0 on Print Bed
Ensure the following:

Figure 2: Critical part dimensions
For each batch printed, the following items must be logged:

Figure 3: Front view of helmet

Figure 4: Strap inserted into buckle

Figure 5: Completed sewn strap